Measuring Magnetic Fields at Cryogenic Temperatures with the MT2000
2023-02-07 · FAQ
Accurate magnetic field measurement is challenging in any environment; at cryogenic temperatures a series of additional considerations come into play.
Hall sensors at low temperature
Of the various methods available for measuring magnetic fields, the simple Hall sensor remains the most versatile even at low temperature. The challenge is to correct the way its performance figures change with temperature, while avoiding unnecessary power dissipation into a very cold environment.
Temperature dependence of Hall sensitivity
Hall sensitivity — the voltage produced for a given field, VH/B — changes with the temperature of the sensor. For typical sensors the average change is around 1 % to 2 % of sensitivity per degree. These changes affect accuracy, but as long as the temperature value is supplied to the instrument, the gaussmeter compensates for the offset automatically.
Quantum oscillations
At both high and low temperature, additional quantum-mechanical behaviour of the 2DEG material becomes noticeable. The Shubnikov–de Haas (SdH) effect causes the Hall sensitivity to oscillate, which affects measurement accuracy. The period of these oscillations varies slightly from sensor to sensor, so each sensor has its own characteristic. What matters in practice is to be aware of the additional measurement uncertainty that appears under these conditions.
Self-heating
A conventional Hall sensor is driven with a current and the resulting Hall voltage is measured. Maximising the drive current maximises the Hall voltage in proportion, which makes the signal easier to measure — but that current dissipates power in the resistive sensor. The result is a balancing act: raise the drive current as far as possible without heating the sensor enough to introduce a temperature error. This matters especially in cryogenic systems, where any heat leaking into the system limits the base temperature, so a measurement solution that minimises dissipation is valuable.
External temperature compensation
The gaussmeter can accept an external temperature value and compensate accordingly, using the known sensitivity variation of the 2DEG Hall sensors used with these instruments.
Bear in mind that this compensates the average sensitivity change only. Because of the unpredictable nature of quantum effects, some additional uncertainty remains at very high fields and low temperatures. That uncertainty is particularly troublesome when accurate incremental measurements are required, because the error can change from positive to negative over a relatively small change of field value.
Cryogenic sensors and probes
Where the field can be accessed from outside — a dewar, for example — a cryogenic Hall probe is a convenient choice. These probes are designed to withstand immersion in cryogen and are very long (up to 150 cm), which allows complete immersion while the probe body is still held outside.
Where access is not possible, or where more accurate positioning is needed, a 2DEG plug-and-play sensor can be mounted on a convenient surface. Installing the sensor and routing the leads out of the environment to the gaussmeter takes more thought and effort, but these sensors offer all the advantages of a calibrated Hall probe together with the flexibility of mounting inside a cryogenic environment.
An in-line connector provides the transition from the sensor wiring to the cable outside the cryogenic environment. The connector can be removed if necessary and the leads soldered to a feedthrough for a specific cryostat. Where a different wiring arrangement is preferred, sensors can be supplied customised to suit.