Short answer: surface field inspection cannot replace coercivity measurement

Surface field inspection measures the field at a point on a magnet’s surface — it reflects the result after magnetizing. BH curve measurement characterizes the material’s demagnetization behaviour and returns remanence Br, induction coercivity Hcb, intrinsic coercivity Hcj and maximum energy product BHmax. The first suits 100% inline screening; the second suits incoming inspection, grade confirmation and demagnetization-resistance analysis. One cannot be converted into the other.

A common mistake is to conclude that coercivity is low because surface field is low. Surface field depends strongly on geometry, size, probe distance and magnetizing fixtures; for the same grade, geometry alone can cause a wider spread than the batch-to-batch spread. When demagnetization resistance is the question, measure the BH curve.

Three methods and when to use them

Method Parameters Sample Notes
DC closed-loop permeameter Br, Hcb, Hcj, BHmax, full demagnetization curve A/B standard samples, small magnets, motor magnet slices Closed magnetic circuit, efficient field use, repeatable; incoming inspection and laboratory work
Pulsed-field BH measurement Demagnetization curve and main magnetic parameters High-coercivity NdFeB, full magnets Very high peak field and short cycle; temperature and system errors need control
Helmholtz coil / fluxmeter Total flux and magnetic moment — not Hcj directly Full magnets and assemblies Fast sorting, not a substitute for coercivity values

The DKP series DC permeameter belongs to the first group (material-level BH and coercivity work). For pole distribution and consistency on the line, use a surface field inspection system; for complete motor stations, see motor test systems.

Reading the four key parameters

  • Br (remanence): flux density remaining after the external field is removed — how much flux the magnet can supply.
  • Hcb (induction coercivity): reverse field required to reduce B to zero.
  • Hcj (intrinsic coercivity): reverse field required to reduce M to zero. This — not surface field — indicates resistance to demagnetization in motor magnets.
  • BHmax (maximum energy product): maximum B×H on the demagnetization curve, a measure of stored magnetic energy.

EV traction magnets are usually evaluated by how much Hcj drops at elevated temperature. For high-temperature coercivity testing, heat the sample and fixture together to the target temperature, allow it to stabilize, then run a full demagnetization–remagnetization cycle. Room-temperature data alone cannot assess hot demagnetization risk.

Where measurement error comes from

  1. Incomplete saturation: if the magnetizing field is too low, the curve does not represent the material and Hcj is underestimated.
  2. Air gap and clamping: gaps between sample and pole faces introduce demagnetizing fields; non-parallel end faces hurt repeatability.
  3. Temperature drift: NdFeB coercivity has a significant temperature coefficient; test conditions must match the application.
  4. Sample geometry: non-standard samples need demagnetizing-factor correction — standard-sample values cannot be reused directly.

Three inputs are enough to select a system: sample size and shape, the parameters required (Br/Hcj only or a full curve), and whether high-temperature testing is needed. Once these are fixed, the permeameter model and fixture design follow.

On the line: 100% surface field screening plus BH sampling

Inline inspection favours cycle time, so it measures peak surface field, pole count, polarity and angle. BH curve measurement takes longer and is normally used for incoming sampling, first-article checks and failure analysis. Linking the two data sets lets surface field act as a fast screen while BH data periodically validates the screening criteria.

To configure a system for your part size, range and cycle time, see our equipment range or contact an engineer for a selection proposal and report template.